pulsarinstruments:psrix_psrix2:psrix2:gain_atten_ctrl [2019/08/12 11:29] ramesh |
pulsarinstruments:psrix_psrix2:psrix2:gain_atten_ctrl [2020/05/08 21:57] (current) ramesh |
||
---|---|---|---|
Line 5: | Line 5: | ||
- connect to the VNC session asterix:4 (use CrabPulsar to connect). | - connect to the VNC session asterix:4 (use CrabPulsar to connect). | ||
- | - load a firmware that has some ADC interfaces. eg. ''python /home/pulsar/roach2/r3xsetup.py'' | + | - load a firmware that has some ADC interfaces. eg. ''python /home/pulsar/roach2/NewFirmware/r3xsetup_new.py'' |
- | - start the ADC displays now: ''python /home/pulsar/roach2/r3xadchist.py'' | + | - start the ADC displays now: ''python /home/pulsar/roach2/NewFirmware/r3xadchist_new.py'' |
- Now change the attenuator setting from another terminal: ''roach2attn <id> <atten>'', where id is 1,2,3 or 4, and attenuation is in the range of 0.5-30.5 dB in steps of 0.5. See example in the screen shot below. | - Now change the attenuator setting from another terminal: ''roach2attn <id> <atten>'', where id is 1,2,3 or 4, and attenuation is in the range of 0.5-30.5 dB in steps of 0.5. See example in the screen shot below. | ||
- Consult the setting on the histogram above - aim to get the signal between -50 to +50. You should see something similar to the plot in the image below. | - Consult the setting on the histogram above - aim to get the signal between -50 to +50. You should see something similar to the plot in the image below. |